China YY9706 Medical Electrical Equipment Electromagnetic Compatibility Testing - China Supplier China YY9706 Medical Electrical Equipment Electromagnetic Compatibility Testing - China Supplier China YY9706 Medical Electrical Equipment Electromagnetic Compatibility Testing - China Supplier China YY9706 Medical Electrical Equipment Electromagnetic Compatibility Testing - China Supplier China YY9706 Medical Electrical Equipment Electromagnetic Compatibility Testing - China Supplier

YY9706 Medical Electrical Equipment Electromagnetic Compatibility Testing

Price:Negotiable
Industry Category: Chemicals/Chemical Intermediates/Intermediate Compounds
Product Category:
Brand:
Spec:


Contact Info
  • Add:廣安門(mén)外大街甲397號(hào), Zip: 100055
  • Contact: 張建忠
  • Tel:13911260297
  • Email:bjkkx2010@163.com

Other Products

Description
Additional Information

Beijing Yizongsuo EMC Testing Center (Institute of Instrumentation Technology and Economy of Machinery Industry, Measurement Control Equipment and Systems Laboratory / Machinery Industry Measurement Control Equipment and Network Quality Testing Center) holds qualifications for electromagnetic compatibility testing of medical electrical equipment, including CNAS and CMA accreditations, and issues formal test reports.

Electromagnetic compatibility testing for medical electrical equipment is conducted in accordance with the YY 9706.102-2021 standard (equivalent to IEC 60601-1-2:2014), which is a specialized standard for electromagnetic compatibility (EMC) of medical electrical equipment, specifying test items and methods. The following is an overview of its main electromagnetic compatibility test items and methods:

I. Electromagnetic Emission (EMI) Test Items

  1. Conducted Emission (CE)
    • Standard Reference: GB/T 6113.201 (CISPR 11)
    • Frequency Range: 150 kHz ~ 30 MHz
    • Method: Measure disturbance voltage at power terminals using an Artificial Mains Network (LISN).
  2. Radiated Emission (RE)
    • Standard Reference: GB/T 6113.201 (CISPR 11)
    • Frequency Range: 30 MHz ~ 1 GHz (if equipment frequency > 1 GHz, extend to 6 GHz)
    • Method: Measure radiated disturbance field strength of equipment and cables using an antenna and receiver in an anechoic chamber.
  3. Harmonic Current Emission
    • Standard Reference: GB 17625.1 (IEC 61000-3-2)
    • Method: Measure harmonic components of the input current (e.g., 2nd to 40th harmonics).
  4. Voltage Fluctuations and Flicker
    • Standard Reference: GB 17625.2 (IEC 61000-3-3)
    • Method: Evaluate the impact of equipment on voltage fluctuations and flicker in the power grid.

II. Immunity (EMS) Test Items

  1. Electrostatic Discharge Immunity (ESD)
    • Standard Reference: GB/T 17626.2 (IEC 61000-4-2)
    • Test Level: Contact discharge ±6 kV, air discharge ±8 kV
    • Method: Apply electrostatic pulses to the equipment enclosure and operating parts to verify immunity.
  2. Radiated RF Immunity (RS)
    • Standard Reference: GB/T 17626.3 (IEC 61000-4-3)
    • Frequency Range: 80 MHz ~ 2.7 GHz
    • Field Strength: 3 V/m (10 V/m for life-support equipment)
    • Method: Radiate an RF field using an antenna in an anechoic chamber to assess whether equipment performance is affected.
  3. Conducted RF Immunity (CS)
    • Standard Reference: GB/T 17626.6 (IEC 61000-4-6)
    • Frequency Range: 150 kHz ~ 80 MHz
    • Level: 3 V (unmodulated)
    • Method: Inject disturbance signals into power lines/signal lines via a coupling clamp or CDN.
  4. Electrical Fast Transient/Burst Immunity (EFT/Burst)
    • Standard Reference: GB/T 17626.4 (IEC 61000-4-4)
    • Test Level: ±2 kV for power ports, ±1 kV for signal ports
    • Method: Apply fast transient bursts to power and signal lines.
  5. Surge Immunity (Surge)
    • Standard Reference: GB/T 17626.5 (IEC 61000-4-5)
    • Test Level: Line-to-line ±1 kV, line-to-ground ±2 kV
    • Method: Simulate lightning or switching surges impacting power ports.
  6. Voltage Dips and Interruptions Immunity
    • Standard Reference: GB/T 17626.11 (IEC 61000-4-11)
    • Test Conditions: Voltage dips to 70%/0% (duration cycles optional)
    • Method: Create voltage dips or interruptions using a power simulator.
  7. Power Frequency Magnetic Field Immunity
    • Standard Reference: GB/T 17626.8 (IEC 61000-4-8)
    • Field Strength: 3 A/m (or higher, e.g., for life-support equipment)
    • Method: Place equipment in a power frequency magnetic field to assess immunity.

III. Special Requirements

  • Life-Support Equipment: Must meet more stringent test levels (e.g., radiated immunity 10 V/m).
  • Performance Criteria:
    • Class A: Normal performance during test;
    • Class B: Temporary performance degradation with self-recovery;
    • Class C: Requires operator intervention to recover (but must not compromise safety).

IV. Test Environment

  • Emission tests must be conducted in an anechoic chamber or open area;
  • Immunity tests must be performed under typical use configurations (including cable layout and accessories).

This standard ensures the safe operation of medical equipment in electromagnetic environments while minimizing interference with other devices. Specific test parameters may be adjusted based on equipment type and intended use environment.

Beijing Yizongsuo EMC Testing Center (Institute of Instrumentation Technology and Economy of Machinery Industry, Measurement Control Equipment and Systems Laboratory / Machinery Industry Measurement Control Equipment and Network Quality Testing Center) is located at No. 397A, Guang'anmenwai Street, Xicheng District, Beijing. It is equipped with a 3-meter anechoic chamber, electrostatic test instruments, surge and burst test instruments, and a full range of EMC testing equipment, as well as capabilities for high/low temperature testing and IP68/IP69 protection level testing.

Industry Category Chemicals/Chemical Intermediates/Intermediate Compounds
Product Category
Brand:
Spec:
Stock: 99999
Manufacturer:
Origin: China / Beijing / Xuanwuqu
About Toocle.com - Partner Programme - Old Version
Copyright ? Toocle.com. All Rights Reserved.
(浙)-經(jīng)營(yíng)性-2023-0192
ChatGlobal Chat Now 清水县| 桂东县| 菏泽市| 菏泽市| 分宜县| 青阳县| 开化县| 临湘市| 志丹县| 木兰县| 赫章县| 浑源县| 洪洞县| 台州市| 宝清县| 偃师市| 巩义市| 湖州市| 竹山县| 密山市| 广灵县| 灵台县| 长寿区| 新余市| 牙克石市| 武功县| 简阳市| 林州市| 奉贤区| 平塘县| 滁州市| 佛坪县| 永泰县| 神木县| 孝昌县| 玉田县| 天台县| 北流市| 绥中县| 万安县| 且末县|